Accurate measurements of voltage and current from electrosurgery are the basis of development of electrosurgery with feedback function. We, therefore, developed a parameter measurement system based on PC, with high voltage and current from electrosurgery being sensed with transformers, amplified, filtered, transformed into single-ended signals, and then into RMS signals. The root mean square (RMS) signals were transformed into digital signals through DAQ card and the data was processed in PC with Labview. The process included sampling, displaying and storage. The experiment results indicated that the measurement system could measure the output parameters from electrosurgery steadily and correctly so that the development of the system has been successful. It can be the basis of development of embedded parameters measurement system and can provide accurate feedback information for intellectual electrosurgery.
Citation:
ZHOUYu, LIDianli, XUWendong, SONGChengli. A New Parameter Measurement System for Electrosurgery Output. Journal of Biomedical Engineering, 2014, 31(2): 421-425. doi: 10.7507/1001-5515.20140078
Copy
Copyright © the editorial department of Journal of Biomedical Engineering of West China Medical Publisher. All rights reserved
1. |
FRIEDRICHS D A,ERICKSON R W,GILBERT J.A new dual current-mode controller improves power regulation in electrosurgical generators[J].IEEE Transactions on Biomedical Circuits and Systems,2012,6(1):39-44.
|
2. |
GOBLE C C O.Electrosurgery system and method:United States,6228081[P].2001-05-08.
|
3. |
ROBY K,VARJAN S,STASCAVAGE K,et al.A novel electrocautery device to increase coagulation rate and reduce thermal damage[C]//2011 IEEE 37th Annual Northeast Bioengineering Conference (NEBEC).Troy,NY:2011:1-2.
|
4. |
DODDE R E,GEE J S,GEIGER J D,et al.Monopolar electrosurgical thermal management for minimizing tissue damage[J].IEEE Transactions on Biomedical Engineering,2012,59(1):167-173.
|
5. |
PICHE N,BARRIERAS D.Minimally invasive nephron-sparing surgery for unilateral Wilms tumor[J].J Pediatr Surg,2012,47(7):e1-e4.
|
6. |
BALAGUER J M,UMAKANTHAN R,LEACCHE M,et al.Minimally invasive cardiac surgery[J].Current Problems in Surgery,2012,49(9):529-549.
|
7. |
WHAM R H,STURM T A.Automatic control system for an electrosurgical generator:United States,7300435[P].2007-11-27.
|
8. |
GINES D L.Electrosurgical generator with adaptive power control:United States,RE40388[P].2008-06-17.
|
9. |
KOSS T,TAIMISTO M H,VARNER R.Impedance mediated control of power delivery for electrosurgery:World Intellectual Property Organization,WO/2011/119933[P].2011-09-29.
|
10. |
ANALOG.High Precision,Wideband RMS-to-DC Converter[EB/OL].(2011-02)[2012-8-20].http://www.analog.com/en/special-linear-functions/rms-to-dc-converters/ad637/products/product.html.
|
11. |
National Instruments.Low-Cost M Series Multifunction Data Acquisition[EB/OL].(2011-10-21)[2012-8-20].http://sine.ni.com/ds/app/doc/p/id/ds-15/lang/zhs.
|
12. |
李刚,张丽君,林凌,等.基于过采样技术的生物电信号检测[J].电子学报,2008,36(7):1465-1467.
|
- 1. FRIEDRICHS D A,ERICKSON R W,GILBERT J.A new dual current-mode controller improves power regulation in electrosurgical generators[J].IEEE Transactions on Biomedical Circuits and Systems,2012,6(1):39-44.
- 2. GOBLE C C O.Electrosurgery system and method:United States,6228081[P].2001-05-08.
- 3. ROBY K,VARJAN S,STASCAVAGE K,et al.A novel electrocautery device to increase coagulation rate and reduce thermal damage[C]//2011 IEEE 37th Annual Northeast Bioengineering Conference (NEBEC).Troy,NY:2011:1-2.
- 4. DODDE R E,GEE J S,GEIGER J D,et al.Monopolar electrosurgical thermal management for minimizing tissue damage[J].IEEE Transactions on Biomedical Engineering,2012,59(1):167-173.
- 5. PICHE N,BARRIERAS D.Minimally invasive nephron-sparing surgery for unilateral Wilms tumor[J].J Pediatr Surg,2012,47(7):e1-e4.
- 6. BALAGUER J M,UMAKANTHAN R,LEACCHE M,et al.Minimally invasive cardiac surgery[J].Current Problems in Surgery,2012,49(9):529-549.
- 7. WHAM R H,STURM T A.Automatic control system for an electrosurgical generator:United States,7300435[P].2007-11-27.
- 8. GINES D L.Electrosurgical generator with adaptive power control:United States,RE40388[P].2008-06-17.
- 9. KOSS T,TAIMISTO M H,VARNER R.Impedance mediated control of power delivery for electrosurgery:World Intellectual Property Organization,WO/2011/119933[P].2011-09-29.
- 10. ANALOG.High Precision,Wideband RMS-to-DC Converter[EB/OL].(2011-02)[2012-8-20].http://www.analog.com/en/special-linear-functions/rms-to-dc-converters/ad637/products/product.html.
- 11. National Instruments.Low-Cost M Series Multifunction Data Acquisition[EB/OL].(2011-10-21)[2012-8-20].http://sine.ni.com/ds/app/doc/p/id/ds-15/lang/zhs.
- 12. 李刚,张丽君,林凌,等.基于过采样技术的生物电信号检测[J].电子学报,2008,36(7):1465-1467.