To address the challenges of capturing micro-strains in detecting esophageal motility disorders and the limitations of existing high-resolution manometry and functional intraluminal imaging probes in directly measuring esophageal tissue electrical impedance, this study proposes a novel flexible balloon sensor structure that integrates a piezoelectric film assembly with a distributed impedance electrode array. Using the electrical analysis module in the finite element analysis (FEA) software, simulations of the forward problem for esophageal impedance detection were conducted to optimize the excitation source parameters, and a physical prototype was fabricated. Under a relative excitation mode with a voltage sensitivity of 2.059%, the voltage output characteristics of the impedance electrode array were analyzed during linear changes in the balloon filling volume. Based on the performance variation of the piezoelectric film assembly, 80% was selected as the optimal filling volume. Force-electric coupling tests were conducted on the balloon sensor using a pressure testing platform, revealing that both the piezoelectric film assembly inside the balloon and the impedance electrodes outside the balloon exhibited significant load differentiation characteristics as the force application point shifted. In summary, this balloon sensor facilitates the localization of force application while simultaneously analyzing esophageal tissue properties, offering a novel diagnostic approach and objective tool for esophageal disease detection.

Copyright © the editorial department of Journal of Biomedical Engineering of West China Medical Publisher. All rights reserved